  • Gastech Air Sampling Pump GSP-300FT-2

    GSP series Air Sampling Pump GSP-300FT-2




    GASTEC developed a detector tube system utilising an electric suction pump in response to growing user needs for improvements in measuring time and routine in the working environment. Moreover, although the reagent filling density of detector tubes is generally high, making effective sampling with conventional electric suction pumps next to impossible at times, the compact GASTEC GSP-300FT-2 has a powerful suction performance that enables effective sampling with any of the designated detector tubes.
    GASTEC's new detection tube measurement system provides performance meeting the relevant measurement criteria for the respective working environment


    Features
    •Flow volume of 50-250mL/min can be set for the inserted detector tube
    •The fixed flow function automatically regulates load changes
    •Real-time flow volume and integrated flow volume are automatically corrected to the values for 20ºC
    •Automatic start enables sampling to begin after a preset time
    •Automatic stop at preset time/volume, with total flow and transpired time calculations
    •Ventilation system configured for uninterrupted sampling
    •Compact and lightweight; 10 hrs. continuous operation on 2 AA batteries
    •Specially-suited detector tubes for continuous workplace sampling (10 min.)
    •Formaldehyde tubes 91P, 91PL can be used
    •Activated carbon and other solid-collecting detector tubes can be utilised

    Specifications

    Integrated flow rate
    0-250mL/min
    Real-time flow rate Parameters
    50mL/min: 0.0-34.5kPa
    100mL/min: 0.0-30.0kPa
    150mL/min: 0.0-25.0kPa
    200mL/min: 0.0-21.0kPa
    250mL/min: 0.0-16.5kPa
    Selectable modes
    Timer mode: Automatic shut-off of air pump at preset time; digital display of integrated flow rate
    Fixed parameter mode: Automatic shut-off of air pump at preset volume; digital display of elapsed sampling time
    Display
    LC Digital Display (backlict)
    Real-time Flow Rate
    mL/min
    Integrated Flow Rate
    0.000-9.999L
    10.00-99.99L
    100.0-999.9L
    Elapsed sampling time
    0-999min
    Configuration / features
    Mass flow sensor (with temperature correction) and diaphragm suction pump, dustproof, splashproof construction, automatic start function, and with initial presetting circuit for maintaining a constant volume of flow
    Accuracy of real-time/integrated flow volume values
    Standard deviation of +/-5% (for a minimum real-time flow of 50mL/min at a fixed temperature)
    Operating Temperature / Humidity range
    0-40ºC(32- 104 ºF) 10-90%RH (no condensation)
    Power source
    Alkaline dry battery AA type (x 2) allows 10 hrs. continuous operation with 2 AA batteries
    Dimensions
    80(W) x 40(D) x 140(H)mm
    Weight
    300g(including batteries)
    Accessories

    Two alkaline dry battery AA type (x 2), detector tube adaptors, dustproof filters (x 5), screwdriver, black suction nipples (already attached before shipping), instruction manual, etc